An HPLC chromatogram is a graph of detector signal over time: each peak is a component leaving the column, and the main peak's share of the total peak area becomes the purity figure on the COA. Learn three landmarks (the axes, the main peak and the small bumps around it) and the graph starts to feel less like lab shorthand and more like a useful map of the sample.
In short: the horizontal axis is time (retention time, in minutes) and the vertical axis is UV detector signal. A clean peptide sample shows one tall, sharp main peak on a flat baseline, with few or tiny secondary peaks. The lab integrates the area under every peak, and the main peak's percentage of the total is reported as HPLC purity.
That is the whole idea in a nutshell. The rest of this guide simply adds the detail you need to look at any chromatogram on a certificate and know what you are seeing. If you would like the broader context on which tests answer which questions, our guide to peptide purity testing methods covers the full toolbox.
Let's start with the frame of the graph. In reversed-phase HPLC, the dissolved sample flows through a column packed with a water-repelling material. As the run goes on, the solvent mixture gradually becomes more organic, and components leave the column one by one: those that cling less come off first, stickier ones later. A UV detector at the end of the column records how much light-absorbing material is passing by, moment by moment.
Near the very start of a run you will often see a small dip or bump as unretained material and the sample solvent pass the detector. Labs generally do not count this region as a peptide peak.
Once the axes make sense, the peak you are looking for is easy to spot. The main peak is the target peptide. On a good chromatogram it is tall, narrow and fairly symmetrical. A main peak with a pronounced shoulder or a long tail usually hints that a nearby component is overlapping with it, or that the separation was not ideal for that run.
Retention time is where that peak appears. If the method stays the same (same column, solvents, gradient and temperature), the same peptide shows up at about the same time in each run. That is why retention time can support identity. When a lab runs an authentic reference standard on the same method and the sample's main peak matches it, that is a retention-time identity confirmation. ILS Laboratories uses this approach and lists it on its reports as Identity (HPLC-RTM).
One note: retention times only make sense within the same method. If a peptide shows up at 5 minutes on one lab's system and 10 minutes on another's, that is not a contradiction.
Almost every real chromatogram has a few smaller bumps around the main peak, and you can read them as clues to the sample's makeup. They are other UV-absorbing components. In synthetic peptides they are usually close relatives of the target, created during synthesis, purification or storage. A useful review of related impurities in peptide products describes the common families (D'Hondt et al., PMID 25044089):
Tiny secondary peaks are normal and expected. What matters is their combined area. If two peaks are close to the same size on a single-compound product, that is a sign the sample is not highly pure, unless the product is a blend, where each component should give its own main peak.
Now for the part behind the purity number. The software sets a baseline under each peak, measures the area inside it, and divides each peak's area by the total of all the peaks. Many reports print this as a peak table. The example below is illustrative only, not taken from any specific certificate:
| Peak | Retention time (min) | Area | Area % |
|---|---|---|---|
| 1 | 8.92 | 1,210 | 0.18 |
| 2 (main) | 9.47 | 667,400 | 99.70 |
| 3 | 10.15 | 790 | 0.12 |
| Total | 669,400 | 100.00 |
Here the main peak holds 99.70% of the integrated area, so the reported purity would be 99.70%. Two details shape the result: which peaks are included in the total, and where the baselines are drawn. ILS Laboratories states on its reports that purity comes from RP-HPLC area normalization at 214 nm: the main peak's share among all peptide-related peaks, with non-peptide process-related impurities left out of the calculation.
Our published reports show the full chromatogram and the computed purity rather than a printed peak table. If you would like the table for a particular lot, it is a reasonable thing to ask a lab or supplier for.
If you have ever wondered why a report says 214 nm, here is the short answer. The peptide bond itself absorbs UV light strongly at low wavelengths, around 210 to 220 nm. Detecting there means almost every peptide and peptide-related impurity shows up, whatever its sequence. A technical overview of peptide HPLC methods is given by Mant and colleagues (PMID 18604941).
The flip side is that material without peptide bonds, such as water, salts and counter-ions like TFA or acetate, contributes little or nothing at this wavelength. That is why a high purity figure does not tell you how much of the powder's weight is peptide. Our guide to HPLC purity vs net peptide content explains the difference.
With the theory covered, the easiest way to get comfortable is to look at real report styles. Our lots are tested by two independent US labs, and their chromatograms look a little different:
On a single-compound lot, both layouts should show one dominant peak with a flat baseline on either side. You can open any of them from the COA registry, and look up the report for your own lot on lot verification using the lot number in your order-confirmation email.
By now you have all the individual pieces. This table pulls them together into a quick checklist.
| Reassuring signs | Worth a closer look |
|---|---|
| Full run shown, start to finish | Trace cropped so the baseline or small peaks are out of frame |
| One sharp main peak on a flat baseline | Shoulders, heavy tailing, or two similar-sized peaks on a single-compound product |
| Small secondary peaks consistent with the reported purity | Visible secondary peaks next to a purity of 100.00% |
| Labeled axes and a named method or wavelength | No axis labels, no method, or a graph that looks pasted in |
| A trace unique to that lot | Identical-looking traces on reports for different peptides |
For a full checklist of certificate warning signs, see fake peptide COA red flags, and for every other field on the report, how to read a peptide COA. Across our lab results dataset of 26 COAs, the median HPLC purity is 99.78%, and our practice is not to release any lot that falls below 99%.
Read time along the bottom and detector signal up the side. Find the tallest peak, which is the target compound, and note its retention time. Then look at the baseline and any small peaks around it. The main peak's share of the total peak area is the purity figure reported on the certificate of analysis.
Each peak is a separate UV-absorbing component. In synthetic peptides, small secondary peaks are usually related impurities such as deletion sequences, oxidized forms or diastereomers. A few tiny peaks are normal. Large secondary peaks on a single-compound product suggest low purity, while blends are expected to show one main peak per component.
By area normalization. The software integrates the area under each peak, adds them up, and divides the main peak's area by the total. If the main peak holds 99.70% of the integrated area, the reported purity is 99.70%. Which peaks are counted, and how baselines are drawn, both affect the result.
Retention time is how long a component takes to travel through the column and reach the detector, measured from the start of the run. Under the same method a given peptide elutes at a reproducible time, so matching a sample's retention time to a reference standard is a recognized way to support identity.
Labs use different columns, solvent gradients, run lengths and detectors, so the same peptide can elute at a different time and the graph can be scaled differently. That is normal. Compare retention times only within one lab's method, and judge each trace on its own: one dominant peak, a steady baseline, and small secondary peaks.
How to Read a Peptide Certificate of Analysis (COA)HPLC Purity vs Net Peptide Content: Why 99% Pure Isn't 100% PeptidePeptide Purity Testing Methods: HPLC, LC-MS and the Rest of the ToolboxFake Peptide COA Red Flags: 12 Signs a Certificate Was Edited, Reused or InventedJanoshik vs Finnrick vs Other Peptide Testing Labs, ExplainedLab results dataset