Peptide purity is usually tested by reversed-phase HPLC with UV detection near 214 nm, while identity is confirmed separately by mass spectrometry or by matching a reference standard. If you have ever looked at a COA and wondered why it lists so many methods, the answer is simpler than it looks: each one answers a single, clear question.
In short: most labs measure peptide purity with reversed-phase HPLC and a UV detector set near 214 nm, reporting the main peak's share of the total peak area. Purity is only one of four questions, though. Identity, quantity and contaminants each need their own method, and a good COA tells you which ones were run.
It helps to picture a lab test as a single, well-defined question. Here are the four that matter most for a lyophilized research peptide:
No single instrument can answer all four questions at once, and that is perfectly normal. A clear COA simply shows which tests were run. Let's start with the method you will see on almost every peptide certificate.
Reversed-phase high-performance liquid chromatography (RP-HPLC) pushes a dissolved sample through a packed column while the solvent mix slowly changes. Each component leaves the column at its own characteristic moment, and a UV detector records it as a peak. Peptide bonds absorb strongly in the low-UV range, which is why 214 to 220 nm is the usual setting. A detailed technical overview of these separation modes is given by Mant and colleagues (PMID 18604941).
The purity number is simply the main peak's area expressed as a share of all the integrated peaks added together, a calculation known as area normalization. Our ILS Laboratories reports spell this out, stating that purity is determined by RP-HPLC area normalization at 214 nm, relative to all peptide-related peaks.
HPLC is excellent at spotting the usual by-products of solid-phase synthesis, such as deletion sequences, truncated chains and oxidized forms. A review of related impurities in peptide products catalogs these families in detail (D'Hondt et al., PMID 25044089). Like any tool, though, it has blind spots:
So HPLC tells you how clean the peptide material is, but not which molecule it is. That is where the next method steps in.
Liquid chromatography paired with mass spectrometry (LC-MS) adds a second detector after the column. As the main peak comes off, the mass spectrometer measures the mass-to-charge ratio of its ions. Peptide ions often carry more than one charge, so you may see a singly charged form labeled [M+H]+ and a doubly charged form labeled [M+2H]2+ at about half that value. When those signals match the theoretical mass of the expected sequence, that is strong evidence the molecule is the right one.
Our Freedom Diagnostics reports work this way. Each one shows a chromatogram, a mass-confirmation spectrum, an LC-MS identity result, an HPLC-UV purity figure and a net content in milligrams.
Mass spec isn't the only legitimate way to answer the identity question. There is a second, perfectly valid approach to confirm identity: run the sample and an authentic reference standard on the same HPLC method and check that their main peaks elute at the same retention time. ILS Laboratories uses this approach, listed on its reports as Identity (HPLC-RTM), confirmed by chromatographic retention-time comparison with a reference standard.
The two approaches answer the question with different kinds of evidence:
Neither is a trick, and both are common. A helpful habit is to check which method a report used, rather than assuming every "identity confirmed" line means a mass measurement was run. Pharmacopeial scientists describe combining several orthogonal techniques, including mass spectrometry, NMR and chromatography, when characterizing peptide reference standards (McCarthy et al., PMID 36949371).
Purity and identity are two big questions, but researchers usually have one more: how much is actually in the vial? Purity is a ratio; quantity is a weight. To learn how many milligrams of peptide are in a vial, labs use a calibrated method such as quantitative HPLC against a reference standard, or amino acid analysis, where the peptide is broken down into its amino acids and each one is measured. Labs that produce reference materials sometimes pair amino acid analysis with quantitative NMR to pin down an especially precise content value (Melanson et al., PMID 30143839).
On our Freedom Diagnostics reports, net content is listed in milligrams for each lot, and per component for blends. The ILS reports we publish also include a net peptide content figure, shown as "report only." Amino acid analysis is not a method we currently run, so it will not appear on our COAs. For why purity and content are different numbers, see HPLC purity vs net peptide content.
So far we have focused on the peptide itself, but a lyophilized powder holds more than peptide, and this is where many people have their "aha" moment. A lyophilized peptide is a powder made of the peptide itself, plus a counter-ion paired with its charged groups, plus a little residual water. Counter-ions such as trifluoroacetate (TFA) or acetate come from synthesis and purification, and they can be a surprisingly large share of the weight. In one Canadian reference-material study, the TFA counter-ion made up nearly 25% of the mass of a peptide sample (PMID 30143839).
Specialized tests measure these components: Karl Fischer titration for water, ion chromatography for counter-ions (Mrozik et al. compare several techniques, PMID 22252914), and gas chromatography for residual solvents. They are routine in drug manufacturing and rare on research-material certificates. Our current COAs don't report Karl Fischer, counter-ion or residual-solvent results, and they don't state a salt form. That is useful to know when you compare certificates from different sources. Our guide to peptide salt forms covers what that means in practice.
Once the peptide chemistry is covered, it is worth asking what else might be in the powder. The limulus amebocyte lysate (LAL) test checks for bacterial endotoxin. Sterility testing looks for live microorganisms, and heavy-metal panels look for metal contaminants. In research settings these mainly matter for cell-based and other sensitive in vitro assays, where trace contaminants can confound an experiment.
Endotoxin is not part of our standard panel, but four of our Freedom Diagnostics reports include an LAL endotoxin assay performed per USP <85>, reported as two passing replicates. Our ILS reports include a fentanyl immunoassay screen with a 50 ng/mL cutoff. We have not run sterility or heavy-metal testing, and a COA should never be read as covering a test it does not list.
If the acronyms are starting to blur together, here is the whole toolbox on one page.
| Method | Question it answers | What it can miss | On NovaForge COAs? |
|---|---|---|---|
| RP-HPLC (UV, ~214 nm) | Purity of the peptide material | Co-eluting impurities; water, salts, counter-ions; identity | Yes, every lot |
| LC-MS | Identity by molecular mass | Isomers with the same mass; quantity | Freedom Diagnostics lots |
| HPLC retention-time match | Identity against a reference standard | Depends on standard quality; not a mass measurement | ILS Laboratories lots |
| Net peptide content (mg) | Quantity in the vial | Purity and identity on their own | Freedom (reported); ILS (report-only figure) |
| Amino acid analysis | Composition and quantity | Sequence order; related impurities | No |
| Karl Fischer | Water content | Everything except water | No |
| Ion chromatography | Counter-ion (TFA, acetate, chloride) content | Peptide identity and purity | No |
| LAL endotoxin | Bacterial endotoxin | Viable organisms; chemical purity | Four Freedom lots |
| Sterility, heavy metals | Microbes; elemental contaminants | Peptide chemistry | No |
Next time a certificate lands in your inbox, try reading it as a list of questions asked and answered:
Every NovaForge lot is tested by an independent US lab before release, and lots below 99% HPLC purity are not released. Your lot number is in your order-confirmation email; enter it on lot verification to pull up the matching report, or browse the COA registry and the lab results dataset. For a field-by-field walkthrough, see how to read a peptide COA, and for a comparison of testing sources, peptide testing labs explained.
Usually by reversed-phase HPLC with UV detection around 214 to 220 nm. The lab integrates every peak in the chromatogram and reports the main peak's share of the total area as the purity percentage. It describes how clean the peptide material is, not which molecule it is or how much is in the vial.
HPLC separates the components of a sample and, with a UV detector, measures how much of the signal belongs to the main peak, which gives purity. LC-MS adds a mass spectrometer after the column, so the lab can also measure the molecular mass of what elutes and compare it with the expected sequence, which confirms identity.
Not by purity alone. HPLC can support identity when the sample's main peak elutes at the same retention time as an authentic reference standard run under identical conditions. That retention-time match is a recognized identity method. Mass spectrometry confirms identity a different way, by measuring the molecule's mass directly.
Amino acid analysis breaks a peptide down into its individual amino acids and measures each one. It is mainly used to confirm composition and to assign an accurate peptide content, especially for reference materials. It does not show the order of the amino acids or separate closely related impurities, so it complements HPLC rather than replacing it.
Because they measure different things. Purity is the main peak's share of the UV-absorbing peptide material. The powder also contains counter-ions such as TFA or acetate and some residual water, which HPLC barely sees but which add weight. Net content counts only the peptide, so it is often well below the total powder weight.
At minimum, a lot-specific identity result with its method, an HPLC purity figure, the lab's name, a report ID and dates. A quantity figure in mg adds useful information. Water, counter-ion, endotoxin and sterility tests are less common on research-material COAs; if a test is not listed, assume it was not run.
HPLC Purity vs Net Peptide Content: Why 99% Pure Isn't 100% PeptideHow to Read a Peptide Certificate of Analysis (COA)Janoshik vs Finnrick vs Other Peptide Testing Labs, ExplainedThird-Party Tested Research PeptidesResearch Peptide GlossaryLab results dataset